Public COA audit

GHK-Cu · KPV

Audited September 5, 2026

Mostly consistent, with concerns

Several fields check out, but we found issues worth reviewing before you rely on this document.

Legitimacy
72
Confidence
85%
Panels
2/7
Purity (printed)
99.789%

Scored moderately because the document is mostly consistent but carries soft concerns, limited panel coverage (2/7), 3 soft concerns, and notable note: "High purity confirmed". To improve: include identity, purity, and endotoxin at minimum.

Document details

Fields extracted from the certificate — not order or payment data.

Batch / lot
665394
Identity method
HPLC with UV Detection Coupled with Mass Spectrometry (LCMS/MS)
Sample date
2026-04-06
Issue date
2026-04-07

Original certificate

Embedded copy of the stored COA PDF or image. Download it for your records — PeptIQ’s audit is separate from the lab document itself.

Loading original certificate…

7-panel checklist

Had data + good, had data + bad, reported without pass/fail, or missing. Missing lowers thoroughness — you decide whether that panel matters; PeptIQ’s score still factors coverage.

2 of 7 present · 2 passed · 0 failed · 0 reported · 5 missing

Missing is a thoroughness gap — not automatically bad. PeptIQ’s legitimacy score still weighs panel coverage; you can down-weight panels that don’t matter for your use case.

  • Identity

    Tested · passed · core

    Compound identity confirmed (e.g. HPLC, mass spec).

    GHK-Cu/KPV/BPC-157/Thymosin Beta-4 confirmed

  • Purity

    Tested · passed · core

    Reported purity percentage for the analyte.

    99.789%

  • Heavy metals

    Not on this certificate file

    Lead, arsenic, cadmium, mercury, or similar limits.

    No evidence on this file. Vendors sometimes publish separate purity and endotoxin reports for the same lot — check the brand page before treating this as “never tested.”

  • Microbial

    Not on this certificate file

    Bioburden / microbial limits testing.

    No evidence on this file. Vendors sometimes publish separate purity and endotoxin reports for the same lot — check the brand page before treating this as “never tested.”

  • Endotoxin

    Not on this certificate file · core

    LAL or chromogenic endotoxin assay.

    No evidence on this file. Vendors sometimes publish separate purity and endotoxin reports for the same lot — check the brand page before treating this as “never tested.”

  • Sterility

    Not on this certificate file

    Sterility testing where applicable.

    No evidence on this file. Vendors sometimes publish separate purity and endotoxin reports for the same lot — check the brand page before treating this as “never tested.”

  • Residual solvent / water

    Not on this certificate file

    Residual solvents or water content limits.

    No evidence on this file. Vendors sometimes publish separate purity and endotoxin reports for the same lot — check the brand page before treating this as “never tested.”

How to read this

Had data · good

Panel present with a pass (or clear acceptance).

Had data · bad

Panel present with a fail — a concrete quality signal.

Had data · unclear

Value reported without an explicit pass/fail mark.

Missing on this file

Not on this certificate file. Lowers file thoroughness; a sibling endotoxin/purity report for the same lot may still exist. PeptIQ’s score still factors coverage.

Findings

Evidence-linked observations from this audit.

  • High purity confirmed

    99.789% purity by HPLC-UV is within normal range for research-grade peptides and is supported by chromatogram showing clean separation of four peptide components

  • Mass spec confirmation present

    Mass spectrometry data shows expected m/z values for all four peptides (GHK-Cu, KPV, BPC-157, TB4) with labeled peaks matching theoretical masses

  • Dates are valid

    Received 2026-04-06, reported 2026-04-07 - both dates are before today (2026-09-05) and show reasonable 1-day turnaround

  • Lab verification limited

    Freedom Diagnostics provides a verification URL (FreedomDiagnosticsTesting.com) and search code (Blan2604060227), but independent verification of lab accreditation/credentials not possible from document alone

  • Limited testing scope

    This COA only covers purity and identity testing. No heavy metals, microbial, endotoxin, sterility, or residual solvent/water testing is included in this report. This is common for purity-focused reports but limits comprehensive quality assessment

  • Batch information present

    Lot number 665394 clearly stated with net peptide content breakdown for all four components

  • Minor formatting inconsistencies

    Date format inconsistent (04/06/2026 vs 04/7/2026 - missing leading zero on day). Product photos show vials but no direct lot number visible on vials in image

PeptIQ COA reports analyze document content and publicly verifiable lab signals. They are educational and not a guarantee of product quality, legality, or safety. Always consult a qualified professional before making health decisions. We never publish purchase prices, shipping details, or uploader identity.